26
Manufacturer’s declaration-electromagnetic immunity
The N507
is intended for use in the electromagnetic environment (for home healthcare) specified
below.
The customer or the user of the N507 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment-guidance
(for home healthcare
environment)
Electrostatic
discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±6 kV
Air±2 kV,±4 kV,±8 kV
Floors should be wood,
concrete or ceramic tile.
If floors are co
vered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power
supply lines
+ 1kV for input/
output lines
+ 2kV for power
supply lines
Not applicable
Mains power quality should
be that of a typical home
healthcare environment.
Surge IEC
61000-4-5
+ 0.5kV, +1kV line(s)
to line(s)
+ 0.5kV, +1kV,+ 2kV
line(s) to earth
+ 0.5kV, +1kV line(s)
to line(s)
Not applicable
Mains power quality should
be that of a typical home
healthcare environment.
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30
cycles
Voltage
interruptions:
0 % UT; 250/300
cycle
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30
cycles
Voltage
interruptions:
0 % UT; 250/300
cycle
Mains power quality should
be that of a typical home
healthcare environment. If
the user of the N507 requires
continued operation during
power mains interruptions,
it is recommended that the
N507 be powered from an
uninterruptible power supply
or a battery.