Communications Section
Command
A
C
N
P
O
L
Function
Sets range of analog
display
Sets memory channel
Inputs ID label
Sets probe connector
number for which data
will be displayed
Sets probe connector
number for which data
will be output
Sets presently selected
settings as default set-
tings
Input format
a. In xyLv or T
∆
uvLv display mode:
“A [ range of
∆
x,
∆
y :
range of
∆
Y ]”
b. In analyzer mode, G standard:
“A [ range of
∆
G :
range of R/G, B/G ]”
c. In analyzer mode, R standard:
“A [ range of
∆
R :
range of G/R, B/R ]”
Acceptable range: 0.1 to 99
Number of characters: Up to 3 (in-
cluding decimal point)
Spaces must not be input before or
after data.
“C memory channel number ”
Acceptable range: 00 to 10 (If op-
tional Card inserted, acceptable range
becomes 00 to 99.)
“N memory channel number
[ 10-character ID label ]”
or “N [ 10-character ID label ]”
Acceptable range of memory channel
number: 00 to 10 (If optional Card inserted,
acceptable range becomes 00 to 99.)
When the optional 4-point expansion
board CA-B04 is used
“P probe connector number ”
When the optional 4-point expansion
board CA-B04 is used
“O probe connector number ”
Acceptable range: 0 to 5
If data output for more than one probe
connector is desired, more than one
probe connector number can be input.
For example, “0134” will cause data from
measuring probes connected to probe
connectors 1, 3, and 4 to be output.
For data output from measuring probes
connected to all probe connectors, input
0 as the probe connector number: “00”
“L”
Further information
•
Corresponds to pressing RANGE
key.
•
Corresponds to pressing MEMORY
CH
or
key.
•
If memory channel number is omit-
ted, memory channel presently be-
ing used will be set.
•
ID label will remain in memory even
if CA-100Plus is switched off.
•
Probe connector number to which no
probe is connected cannot be selected.
•
Corresponds to pressing PROBE key.
•
Probe connector number to which no
probe is connected cannot be selected.
•
The presently selected probe connec-
tor number for data output can be
determined by using the command
“Z” (status information request).
•
If probe connector number for
which data will be output is not se-
lected after POWER switch is first
set to ON, the probe connector num-
ber for data output will be the same
as the probe connector number for
which data will be displayed.
•
Changes the default settings (settings
which are automatically set when POWER
is first set to ON) for the following items:
a. Display mode
b. SYNC mode
When the optional 4-point expansion
board CA-B04 is used
c. Memory channel
d. Probe connector for which data
will be displayed.
•
Corresponds to pressing
key
while
key is held pressed.
*1
*2
*1
*2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание CA-100PLUS -
Страница 25: ...22 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 47: ...44 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 101: ...98 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 123: ...120 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 124: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...