Communication Format
The commands listed in the following tables can be used to control the CA-100Plus. Most of the commands listed
are applicable to data communication with RS-232C system; those that are applicable to only one system are
indicated as such.
1. Input Command Table (PC
→
Instrument)
Inputting commands shown in the table below will operate the instrument in the same way as when the correspond-
ing keys are pressed.
Terms used in the table
Display probe no. Indicates the probe connector no. whose measured data is displayed.
Output probe no.
Indicates the probe connector no. whose measured data is to be output.
CAL mode
Indicates that the
key has been pressed.
Input Command Table
Command
I
M
S
H
U
Function
Performs zero calibra-
tion
Selects display mode
Selects SYNC mode
Sets/cancels display
hold
Setting user-selected
calibration data or stan-
dard color data (for
memory channel 00)
Input format
“I”
“M0”: Sets xyLv display mode
“M1”: Sets T
∆
uvLv display mode
“M2”: Sets analyzer (RBG) display
mode
“S0”: Sets NTSC SYNC mode
“S1”: Sets PAL SYNC mode
“S2”: Sets EXT SYNC mode
“S3”: Sets UNIV. SYNC mode
“H0”: Cancels display hold
“H1”: Sets display hold
“U [ x value : y value : Lv value ]”
•
x value or y value can be input in
any of the three formats given in the
examples below for inputting the value
0.330:
“0.330”
“.330”
“330”
Spaces must not be input before or after data.
•
Lv value
Acceptable range: 0.01 to 999
Number of characters: Up to 4 (in-
cluding decimal point)
Spaces must not be input before or
after data.
Further information
•
Block all light from reaching the re-
ceptor area of measuring probes be-
fore inputting this command.
•
Until this command is input and
zero calibration is completed, the
commands U, A, N, J, and E will
not be accepted. (Do not string the
command I together with other
commands using “&”.)
•
Corresponds to pressing 0-CAL key.
•
Corresponds to pressing MODE
key.
•
Corresponds to pressing SYNC
key.
•
Corresponds to pressing HOLD
key.
•
Corresponds to pressing
key in
xyLv or T
∆
uvLv display mode.
•
Cannot be used in analyzer mode.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание CA-100PLUS -
Страница 25: ...22 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 47: ...44 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 101: ...98 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 123: ...120 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 124: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...