74
74
2. Checking the Probe Serial No. when Making Settings
To check the probe serial no. when making settings, press the MR
key for two to four seconds (a bleep will sound after two seconds
have elapsed) and check it in the LCD display section.
<When xyLv, T
∆
uvLv, u'v'Lv or XYZ mode is selected>
The serial number of the probe used when user calibration is performed or when target color is set will be dis-
played.
When the optional 4-point expansion board CA-B04 is used
The probe serial no. of the probe connector used for the current selected memory channel will be displayed.
1
Probe serial no. used when user calibration was performed
2
Probe serial no. used when the target color was set
By default (factory setting), “00000000” is set for both
1
and
2
.
•
When “00000000” is set for
1
: If measurement is performed with this memory channel, Minolta’s cali-
bration standard will be used for the measurement. (Same as when
measurement is performed with the memory channel CH00.)
•
When “00000000” is set for
2
: x=0.3127, y=0.3293 and Lv=160.0 (cd/m
2
) are set as the values of the
target color.
<When an analyzer measurement mode is selected>
The probe serial no. that was used to input the analyzer mode RGB emission characteristic or set the target color
for the currently selected memory channel will be displayed.
When the optional 4-point expansion board CA-B04 is used
The probe serial no. of the probe connector used for the current selected memory channel will be displayed.
1
Probe serial no. used to input the analyzer mode RGB emission characteristic
2
Probe serial no. used when the target color was set
By default (factory setting), “00000000” is set for both
1
and
2
.
•
When “00000000” is set for
1
: The RGB emission characteristic for the display’s analyzer mode has
not been input.
* The serial no. of the currently used measuring probe can be viewed in the PROBE selection screen, that can be opened by pressing the
then
keys.
(If the 4-point expansion board CA-B04 is used, the probe connector no. will switch from one to another each time the
key is pressed.
For details, refer to page 43.)
SELECT : PROBE
P1 35881112
0
2
4
(sec)
Period for which the MR key is
pressed
Bleep
Bleep
Bleep
Target color
is displayed.
The probe serial
no. used to mak-
ing settings is
displayed.
The unit of
intensity is
displayed.
16790160
16790160
16790160
16790160
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание CA-100PLUS -
Страница 25: ...22 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 47: ...44 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 101: ...98 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 123: ...120 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 124: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...