4
Sampling rate
200MSa/s
Waveform
amplitude
resolution
12 bits
The minimum
frequency
resolution
10mHz
Frequency error
±5×10
-6
Frequency stability
±1×10
-6
Amplitude range
(peak to peak)
15mVp-p~15Vp-p(12MHz or less)
15mVp-p~8Vp-p(12MHz above)
Output Impedance
50Ω±10%
Amplitude
resolution
1mVp-p
(
-20dBAttenuation
)
10mVp-p
(
Does not decay
)
Amplitude stability
±0.5%
(
Every five hours
)
Amplitude error
± 1%+10mV
(
Frequency1KHz
,
15 Vp-p
)
Offset Range
-
120%
~
+120%
(
The ratio of the bias voltage and signal
amplitude
)
Bias Resolution
1%
Phase range
0
~
359°
Phase resolution
1°
Sine wave
Harmonic arrived
System
40dBc(<1MHz) ,35dBc(1MHz~20MHz)
Distortion
<0.8%(20Hz~20KHz)
Square wave
Lifting along time
≤20ns
Overshoot
≤10%
Duty cycle
adjustment range
0%~99.9%
TTL
Lifting along time
≤20ns
LOW
<0.3V
High
1V~10V
Arbitrary
Waveform
Quantity
16 groups
Memory depth /
group
1KB / 16 group
Scan
Scan Mode
Linear sweep, log sweep
Scan time
1S
~
999S
Scan range
It is determined by the sweep parameter settings
External
measurements
Frequency range
GATE-TIME=10S 0.1HZ - 60MHZ
GATE-TIME=1S 1HZ - 60MHZ
GATE-TIME=0.1S 10HZ - 60MHZ
GATE-TIME=0.01S 100HZ - 60MHZ