Other Functions
26
5.2 Messages Prompting
The Pulse Oximeter can display various prompt messages. In table 5-1,
prompt messages as well as their causes and solutions are listed.
Table 5-1 Table of Indications
Message
Cause
Solution
“Low Battery”
Batteries energy lower
than 4.0 Voltage.
Replace batteries in time
“Memory Full”
The internal memory
is almost full.
Stored data is to be covered.
Export data in time.
Blinking
“Memory Full”
Memory is full.
Stored data is being
covered. Export data in time.
“ID Full”
ID number Stored is
greater than 95.
ID data is to be covered.
Export data in time.
Blinking “ID Full”
ID data is being
covered.
Export data in time.
“Standby”
The device is in
standby mode.
None.
“Communication”
The device is in
communication mode.
None.
DELETE ALL?
The Delete ID button
is pressed.
Refer to
5.1.4 Data Deletion
ALL DELETED
The Delete ID button
is pressed again after
“DELETE ALL?”
appears.
None.