Safety Information
2
Warning
The Pulse Oximeter is to be operated only by medical doctors or
trained healthcare personnel to measure the SpO
2
of clinical
patients (including adult, pediatric and neonatal patients) in
hospitals and healthcare agencies or in the process of transporting
patients.
The Pulse Oximeter is designed for real-time and rapid
measurement of the patient’s SpO
2
. It is not suitable for long-time
continuous patient monitoring. Continual measurement must not
exceed 2 hours.
Before using the Pulse Oximeter, the user should check and make
sure the device and its accessories can work properly and safely.
When using the Pulse Oximeter together with the electrical surgery
equipment, the user should pay attention to and guarantee safety
of the patient being measured.
To avoid fire or explosion hazards, do not use the Pulse Oximeter
at places where such flammable material as anesthesia gas is
present.
Do not pull or lift the Pulse Oximeter by its connection cable. That
may lead to device falling and consequent patient injuries.
It is not recommended to hang the Pulse Oximeter when
transporting patients. Safety hazards may arise from the large
amplitude swing during the transportation.
Do not use the Pulse Oximeter or its sensor around working MRI
equipment. Induction current may lead to damages. The device
may affect operation of MRI equipment and vice versa.
The Pulse Oximeter and its accessories may be contaminated by
microorganism during transporting, use and storage. Use the