Other Functions
24
The first trend data will be stored 15 seconds after the pulse is found.
Thereafter one trend data will be stored every 2 minutes. One trend
data includes:
Average SpO
2
value within the 2 minutes.
Average PR value within the 2 minutes.
Note
The first trend data are the instant values of SpO
2
and PR.
For a patient, his (her) ID data are calculated from all measured
trend data of him (her).
Once the trend data of a patient is covered, all trend data of that
patient will be deleted, however, the ID data will not be deleted.
5.1.2 Data Adding
The previously stored ID number appears on the screen when a finger
is inserted into the SpO
2
sensor. It will keep blinking for 8 seconds after
the pulse is found.
Press the Confirm ID button before the ID number stops
blinking, the Pulse Oximeter will set the ID number as current
patient ID. The data measured thereafter will be superadded
to the previous ID.
If the user doesn’t press the Confirm ID button before the ID
number stops blinking, a new ID number, which is the blinking
number plus 1, will be set as the current patient ID.
The Confirm ID button is of no use when the current ID number is 000,
which will automatically change to 001 when the pulse is found.