3 - 39
Repair Information
Troubleshooting Menus
3.16.2
NIBP Troubleshooting
ECG Channels Checking...
1
Appears when the ECG
Module is calibrating.
1
This is normal operation.
Turn off ECG module
calibration and the
message will be cleared.
Initialization Error
1
During the ECG module
power-on, as the ECG
module communication
stops, system fails to
communicate with
module.
1
Contact Technical
Support.
MESSAGE/PROBLEM
REASON
SOLUTION
MESSAGE/ PROBLEM
REASON
SOLUTION
NIBP: Self Test Error
1
Failed self-test.
2
Sensor or A/D sampling
may have error.
1
Power cycle unit. If
message reappears,
contact Technical
Support.
NIBP: Communication Error
1
NIBP module
communication error. The
command can not be sent
correctly.
1
Power cycle unit. If
message reappears,
contact Technical
Support.
NIBP: System Error
1
System error. After start-
up, the inflating pump,
A/D sampling unit or
pressure sensor may have
error or the pointer may
have error when the
software is running.
2
Power supply is not stable
or circuit failure leads to
voltage error.
1
Power cycle unit. If
message reappears,
contact Technical
Support.
NIBP: Cuff Overpressure
1
The hardware
overpressure limit has
been exceeded.
2
Overpressure. The cuff
pressure exceeds 297
mmHg in ADU mode,
240 mmHg in PED mode
or 147 mmHg in NEO
mode.
1
Power cycle unit.If
message reappears,
contact Technical
Support.
NIBP: Pneumatic Leak
1
Leakage. In Pneumatic
check, air leakage is
found in hose.
1
Change the cuff.
NIBP: Unable to Measure
1
Unable to make
measurement after three
automatic retries.
1
Check Patient.
2
Retry measurement. If
message reappears,
power cycle unit. If
message reappears,
contact Technical
Support.
NIBP: Reset Failed
1
Reset failed.
1
Power cycle unit.If
message reappears,
contact Technical
Support.