A - 6
WARNING
•
The infusion accuracy and pressure detection is affected by viscosity of
liquids and disposables used (for example diameter, plunger, material, and
needle).
NOTE
•
The infusion accuracy tests and occlusion pressure tests are performed in
accordance with IEC60601-2-24:2012 (test temperature:
20°C ± 2°C
).
A.8
Recommended Syringes
VTBI set range
0.01 to 9999.99 ml
Resolution: 0.01ml
Weight set range
0.1 to 499.0 kg/0.2 to1100.1 lb
Drug Amt. set range
0.001 to 99999
Drug Amt. unit set range
ng, μg, mg, g, mU, U, kU, EU, mmol, mol, mcal, cal, kcal,
mEq
Volume
0.10 to 9999.99ml
Conc. set range
0.001 to 9999.99
Conc. unit set range
ng/ml, μg/ml, mg/ml, g/ml, mU/ml, U/ml, kU/ml, EU/ml,
mmol/ml, mol/ml, mcal/ml, cal/ml, kcal/ml, mEq/ml
Dose Rate set range
0.001 to 99999
Product Name
Size
Manufacturer
Sterile Hypodermic
Syringes for Single Use
1ml, 5ml, 10ml, 20 ml, 30 ml, 50 ml
Double-Dove
B. Braun Original Perfusor
Syringe
20ml, 50ml
B. Braun Melsungen
AG
B. Braun Omnifix Luer Lok
Solo
2ml, 3ml
B. Braun Melsungen
AG
Содержание BeneFusion eSP Vet
Страница 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Страница 2: ......
Страница 12: ...6 This page intentionally left blank ...
Страница 26: ...2 10 This page intentionally left blank ...
Страница 44: ...5 4 This page intentionally left blank ...
Страница 74: ...11 8 This page intentionally left blank ...
Страница 82: ...12 8 This page intentionally left blank ...
Страница 100: ...A 10 This page intentionally left blank ...
Страница 108: ...B 8 This page intentionally left blank ...
Страница 111: ...D 1 D Declaration of Conformity ...
Страница 112: ...D 2 This page intentionally left blank ...
Страница 113: ......
Страница 114: ...P N 046 021887 00 1 0 ...