6 - 3
6.3
Resetting Alarms
When an alarm occurs, press
to acknowledge and reset the alarm. The alarm reset
state has the following features:
■
A
√
appears before the alarm message, indicating that the alarm is
acknowledged.
■
The alarm sound is silenced, and the alarm screen disappears.
■
The alarm reset symbol
is displayed after the alarm message.
For the following alarms, when they are reset, all the alarm indications (alarm sound,
alarm message, and alarm light) disappear.
■
Syringe Empty
■
VTBI Complete
■
KVO Finish
■
Standby Time Expired
■
Extension Line Detached
6.4
Pausing Alarm Sound
To enter the audio pause state, choose one of the following ways:
■
Select
in the alarm screen.
■
Swipe the touchscreen from top down, and select
.
The audio pause state has the following features:
■
Except for the
Battery Depleted
alarm, the sound of all alarms are silenced for two
minutes.
■
The audio pause symbol
is displayed in the system information area.
■
If a new alarm is triggered during the audio pause state, the sound of the new
alarm will also be silenced.
When the audio pause time expires, the audio paused state is automatically deactivated.
You can also cancel the audio paused state by pressing
again.
For the
Low Battery
,
Reminder
,
Time Near End
and
Syringe Near Empty
alarms, press
and the pump gives a reminder sound every 5 minutes. The symbol
is displayed
after the alarm message.
NOTE
•
Except for the Battery Depleted alarm, the sounds of all alarms are paused by
pressing
.
Содержание BeneFusion eSP Vet
Страница 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Страница 2: ......
Страница 12: ...6 This page intentionally left blank ...
Страница 26: ...2 10 This page intentionally left blank ...
Страница 44: ...5 4 This page intentionally left blank ...
Страница 74: ...11 8 This page intentionally left blank ...
Страница 82: ...12 8 This page intentionally left blank ...
Страница 100: ...A 10 This page intentionally left blank ...
Страница 108: ...B 8 This page intentionally left blank ...
Страница 111: ...D 1 D Declaration of Conformity ...
Страница 112: ...D 2 This page intentionally left blank ...
Страница 113: ......
Страница 114: ...P N 046 021887 00 1 0 ...