10 - 1
10
Managing Patient
10.1
Discharging/Admitting a Patient
Before admitting a new patient, discharge the previous patient. After the patient is
discharged, all patient data are removed from the pump. After a patient is discharged,
the pump automatically admit a new patient.
The patient is automatically discharged in the following cases:
■
After the patient data is successfully exported through the USB drive. For more
information, see
10.3 Exporting Patient Information
.
■
After the patient is discharged by the CMS or the patient monitor.
To manually discharge a patient, follow this procedure:
1.
Swipe the touchscreen from top down
→
select
Menu
→
select
Discharge
Patient
.
2.
Select
Accept
.
WARNING
•
Always discharge the previous patient before starting an infusion. Failure to
do so can lead to data being attributed to the wrong patient.
10.2
Editing Patient Information
Edit patient information after a patient has been admitted, or when patient information
is incomplete, or when you want to change patient information.
To edit patient information, follow this procedure:
1.
Swipe the touchscreen from top down
→
select
Menu
→
select
Patient
Management
.
2.
Edit patient information as required.
10.3
Exporting Patient Information
To export the information of the current patient to the USB drive, follow this procedure:
1.
Connect the USB drive to the USB connector.
2.
Swipe the touchscreen from top down
→
select
Menu
→
select
Patient Manage-
ment
→
select
Export Patient Information
.
Содержание BeneFusion eSP Vet
Страница 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Страница 2: ......
Страница 12: ...6 This page intentionally left blank ...
Страница 26: ...2 10 This page intentionally left blank ...
Страница 44: ...5 4 This page intentionally left blank ...
Страница 74: ...11 8 This page intentionally left blank ...
Страница 82: ...12 8 This page intentionally left blank ...
Страница 100: ...A 10 This page intentionally left blank ...
Страница 108: ...B 8 This page intentionally left blank ...
Страница 111: ...D 1 D Declaration of Conformity ...
Страница 112: ...D 2 This page intentionally left blank ...
Страница 113: ......
Страница 114: ...P N 046 021887 00 1 0 ...