A - 7
NOTE
•
The recommended extension set is B.Braun Original Perfusor Line (using IV-
Standard-PE, and with Luer lock).
A.9
Occlusion Alarm Delay and Bolus Volume
Test conditions:
■
Syringe brand: B.Braun Original Perfusor Syringe, B. Braun extension line
■
Test temperature: 20
°C
±2
°C
WARNING
•
Occlusion alarm pressure, alarm delays and bolus volume may vary
depending on test conditions, temperature and tube length. Using syringe of
a larger size and infusing at a lower rate may cause longer occlusion alarm
delay.
Syringe size (ml)
Rate (ml/h)
Occlusion alarm delay time (hh: mm: ss)
High occlusion alarm
pressure level
Low occlusion alarm
pressure level
20
1
<
00:33:12
<
00:02:12
5
<
00:06:54
<
00:00:45
50
1
<
01:45:57
<
00:06:53
5
<
00:25:15
<
00:01:42
Syringe size (ml)
Rate (ml/h)
Bolus volume after occlusion (ml)
High occlusion alarm
pressure level
Low occlusion alarm
pressure level
20
5
<
0.2
<
0.1
50
5
<
0.3
<
0.15
Содержание BeneFusion eSP Vet
Страница 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Страница 2: ......
Страница 12: ...6 This page intentionally left blank ...
Страница 26: ...2 10 This page intentionally left blank ...
Страница 44: ...5 4 This page intentionally left blank ...
Страница 74: ...11 8 This page intentionally left blank ...
Страница 82: ...12 8 This page intentionally left blank ...
Страница 100: ...A 10 This page intentionally left blank ...
Страница 108: ...B 8 This page intentionally left blank ...
Страница 111: ...D 1 D Declaration of Conformity ...
Страница 112: ...D 2 This page intentionally left blank ...
Страница 113: ......
Страница 114: ...P N 046 021887 00 1 0 ...