11.14 The Purge Limit Settings
11.15 The Parameter Memory Setting
11.16 The Loading Guide Setting
11.17 The Brand Selection Setting
11.18 The Auto-Restart Setting
Menu Item
Default Setting
Function
Purge Limit
1ml
Sets the maximum volume of the purge. The
purge stops when the set volume is reached. The
setting range is 0.01 ml to 5 ml.
Menu Item
Default Setting
Function
Para. Memory
Off
Sets the parameter memory switch. If this switch
is turned on, the pump can automatically reload
the infusion mode and other infusion parameters
when restarted if the same drug has been
selected.
Menu Item
Default Setting
Function
Loading Guide
On
Sets whether enter the loading guide screen
when the syringe is not loaded.
Menu Item
Default Setting
Function
Brand Selection
On
Sets whether the brand list will be displayed after
the syringe is loaded or replaced.
Menu Item
Default Setting
Function
Auto-restart
Off
Sets whether to
restart the infusion or not
when the occlusion pressure is reduced.
Содержание BeneFusion eSP Vet
Страница 1: ...BeneFusion eSP Vet Veterinary Syringe Pump Operator s Manual ...
Страница 2: ......
Страница 12: ...6 This page intentionally left blank ...
Страница 26: ...2 10 This page intentionally left blank ...
Страница 44: ...5 4 This page intentionally left blank ...
Страница 74: ...11 8 This page intentionally left blank ...
Страница 82: ...12 8 This page intentionally left blank ...
Страница 100: ...A 10 This page intentionally left blank ...
Страница 108: ...B 8 This page intentionally left blank ...
Страница 111: ...D 1 D Declaration of Conformity ...
Страница 112: ...D 2 This page intentionally left blank ...
Страница 113: ......
Страница 114: ...P N 046 021887 00 1 0 ...