Accutorr Plus™ Operating Instructions
6 - 7
Appendix
Specifications
6.2.11
Physical Characteristics
• Size (maximum):
6.2.12
Environmental Characteristics
• Operating:
Main Unit:
19 cm(W) x 26.93 cm(H) x 20.83 cm (D)
7.5" (W) x 10.6" (H) x 8.2" (D)
Recorder Module:
5.33 cm(W) x 23 cm(H) x 11 cm (D)
2.1" (W) x 9" (H) x 4.25" (D)
Predictive Module:
5.7 cm(W) x 15.9 cm(H) x 11.8 cm (D)
2.25" (W) x 6.25" (H) x 4.63" (D)
Weight:
<4.95 kg (11 pounds), depending on
configuration
Temperature:
10°C to 40°C, 50°F to 104°F (Accutorr Plus
& Recorder)
10°C to 32°C, 50°F to 90°F (Predictive
Thermometer)
Humidity:
15 to 90% max, non-condensing.
Altitude:
1013 hPa to 697 hPa (0 to 10,000 ft.)
Shock and Vibration:
Meets IEC 60068-2-27 for shock with peak
acceleration of 15g, 11 mSec duration, half
sine.
Meets EN 60068-2-64 for random vibration
with frequencies of 10 to 2000 Hz,
resolution of 10Hz, 10 minutes per axis,
acceleration spectral density of:
10 Hz to 100 Hz : 1.0 (m/s2)2/Hz
100 Hz to 200 Hz : -3 dB per octave
200 Hz to 2000 Hz : 0.5 (m/s2)2/Hz
Shipping:
Meets ISTA Test Procedure 1A
(less than 100 lbs)
• Storage:
Temperature:
-15°C to +60°C,
+5°F to 104°F
Humidity:
10 to 95%, non-condensing