Temperature Measurement (Optional)
Operation
3 - 28
Accutorr Plus™ Operating Instructions
3.11.4
Storing Temperature Measurements
Predictive temperature measurements are automatically stored in the trend memory.
AccuTemp IR temperature measurements are stored in the trend memory only if the AccuTemp
I.R. thermometer is returned to the Accutorr Plus within 60 seconds of the reading. Welch
Allyn Sure Temp Plus Thermometer measurements are not stored in the trend memory.
When a temperature measurement is completed within 2 minutes before or after an NIBP
measurement, it is stored as occurring at the same time as the NIBP measurement. If more
than one temperature measure is taken during this ±2 minutes, then only the last temperature
measurement is stored.
When a temperature measurement is taken outside of this ±2 minutes, then it is stored as an
individual item. Also, when temperature measurements are taken within two minutes of each
other, the newer measurement replaces the older measurement. When more than 2 minutes
passes between temperature measurements, then each measurement will be stored.