Accutorr Plus™ Operating Instructions
3 - 29
Operation
Recorder (Optional)
3.12
Recorder (Optional)
The Accutorr Plus can provide a permanent record of patient data using the PRINT key (12).
There are two print modes available. They are Continuous Print or Request Print. In the
Continuous Print mode the printer will print each time there is a valid NIBP or Temperature
measurement. In the Request Print mode the printer will print all of the stored information for
the displayed patient.
1.
Attach the Recorder Module as shown in section 3.17.
2.
Press the
key (12) (1 beep tone) to generate a Request printing. The recorder will
print all stored measurements for the currently displayed patient. Press the
key (1
beep tone) while a printing is in progress, to stop the printing.
3.
Press and hold the
key (12) (2 beep tones, approx. 3 seconds) to switch the print
mode between Continuous and Request. When in the Continuous mode the Print LED
(13) is illuminated.
NOTE:
When a printing is in progress and the PRINT key is pressed
or Room Number and/or Bed Letter is changed, the printing
will stop.
FIGURE 3-11
Recorder Strip Sample
When the Predictive thermometer is used, “P” is printed next to the temperature
measurement. When the IR thermometer is used, “I” is printed next to the temperature
measurement. When NIBP is used to obtain a pulse rate measurement, “N” is printed next to
the pulse rate measurement. When SpO
2
is used to obtain a pulse rate measurement, “S” is
printed next to the pulse rate measurement. If data is not available for any given parameter,
“—-” is printed under that parameter. Parameter values that violated alarm limits are
indicated by the brackets “[ ]”.
M/D/Y 11/25/97 2a
HH:MM SYS DIA MAP
15:25[122] [ 88] [ 99]
BPM SPO2 °F/C
[S 64] [99] P 98.9
HH:MM SYS DIA MAP
15:20 120 [ 88] 99
BPM SPO2 °F/C
S 64 99 P 98.9
HH:MM SYS DIA MAP
15:15 120 88 99
BPM SPO2 °F/C
S 64 99 – – – –
HH:MM SYS DIA MAP
15:10 120 [ 88] 99
BPM SPO2 °F/C
[S 64] 99 P 98.9
HH:MM SYS DIA MAP
15:05 120 [ 120] 88 99
BPM SPO2 °F/C
S 64 99 P 98.9
The Date and Room/Bed is printed
for each group of measurements.
Parameter Headings are repeated
for each line of measurements.
Brackets are printed around measurements
that caused an alarm violation.
P or I is printed with the Temp
measurement, indicating the temperature
was acquired from a Predictive or the
AccuTemp IR thermometer.
When no information is available for a
particular parameter, dashes are printed.
S or N is printed with the Pulse Rate (BPM)
measurement, indicating the Pulse Rate
was acquired from SpO
2
or NIBP.