Setting-up / Turning Power On
Operation
3 - 2
Accutorr Plus™ Operating Instructions
5.
Press the
ON/STANDBY
key (19) to activate the unit. If it is required to enter the User
Configuration mode, press and hold the
DEFAULTS
key (14) while the unit is powering
on. See section 3.15 for more details on the User Configuration mode.
6.
The unit will count down from 20, display all 8’s in the LEDS and perform internal
diagnostic tests. Any status codes are displayed in the appropriate LED. See section
3.16 for a list of status codes. At the end of power up, all of the displays illuminate and
then blank, except the Bed Letter and Room Number displays (25 & 26) which does not
blank. A beep tone will sound during the power up sequence to confirm the operation of
the audio indicator. If the time and date need to be set, see section 3.13 for instructions.
7.
On Accutorr Plus advanced models, adjust the contrast on the LCD if necessary. To
adjust the contrast, press and hold the MENU key (8) (2 beep tones, approx. 3
seconds). Use the LCD Up & LCD Down Arrow keys (9 & 10) to adjust the contrast. See
section 3.8, Setting the LCD Contrast (View Angle Adjustment), for more details.