- 5-30 -
13. Touch the [ENTER] and then the checking
is started.
• When the UV lamp lights off, it takes
time to light.
14. After checking, open the left maintenance
cover.
15. Remove the UV illuminometer BKT 2 and
then verify the level on the indicator.
• Refer to the indicator value inside of
the UV illuminometer.
•
To avoid a burn injury ,do not touch
the left side of the BKT 2 after check-
ing the level.
16. After verifying the level, touch the [EN-
TER].
17. Touch the JOG [ ] to complete the illumi-
nation check.
When checking the illumination again, touch the
JOG [ ] and do the step E of the ‘‘Setting the UV
illuminometer’’.
•
Make sure to check the level again
after the lamp cools down. An accu-
rate result can not be obtained by UV
lamp heat.
UV LEVEL CHECK
CHECK<
>FINISH
PLEASE CLOSE
THE LEFT COVER
UV LEVEL CHECK
CHECK START
:ent
UV LEVEL CHECK
CHK THE TOOL
:ent
PLEASE OPEN
THE LEFT COVER
UV LEVEL CHECK
CHECK START
:ent
** CHECKING **
PLEASE WAIT
Содержание UJF-605RII
Страница 22: ... xx ...
Страница 82: ... 2 40 ...
Страница 104: ... 3 22 ...
Страница 114: ... 4 10 ...
Страница 148: ... 5 34 ...
Страница 162: ... A 6 ...
Страница 168: ...A 12 D201670 1 10 27122007 ...
Страница 169: ......
Страница 170: ...Printed in Japan MIMAKI ENGINEERING Co Ltd 2007 EM FW 1 10 ...