- 22 -
POD Unit, tubing and power cord,
Continued
Loop
•
Install the Inlet Strainer as shown here.
•
Connect one end of the feedwater tubing to the Inlet Strainer.
NOTE:
•
A pressure regulator is normally required after the Inlet Strainer.
Connections to
System Cabinet
Follow the steps below.
Step Action
Diagram
1
Plug one end of the feedwater
tubing to the Cabinet.
Open the valve on the other
end of the feedwater tubing to
allow water flow later.
2
Plug in the Termination Plug.
It must be plugged in
before the power cord.
3
Plug in the power cord. The
Main Display goes through a
series of start up screens.
4
Wait for the Main Display to
show a STANDBY Mode
screen. This may take up to a
few minutes.
Continued on next page
1
2
3