Summary of Alarm messages
Alarm
messages
LCD message
What it means
FLOW AUTO STOP
The System has automatically
stopped dispensing water. The POD
FLOW STOP timer has reached 0
minutes.
Push the POD Unit Plunger all the
way down and release.
This resets the dispenser timer and
makes the POD Unit available for
dispensing.
INCORRECT Q-GARD PACK
The System does not recognise the
type of Q-Gard Pack being installed.
Contact Millipore.
INCORRECT QUANTUM
CARTRIDGE
The System does not recognise the
type of Quantum Cartridge being
installed.
Contact Millipore.
MILLI-Q FEED C > MAX
The feedwater conductivity is out of
measurement range.
Contact Millipore.
MILLI-Q FEED T < MIN
The feedwater temperature is out of
measurement range.
Contact Millipore.
MILLI-Q FEED T > MAX
The feedwater temperature is out of
measurement range.
Contact Millipore.
MILLI-Q INTER R > MAX
The Intermediate resistivity is out of
measurement range.
Contact Millipore.
MILLI-Q INTER T < MIN
The Intermediate temperature is out
of measurement range.
Contact Millipore.
MILLI-Q INTER T > MAX
The Intermediate temperature is out
of measurement range.
Contact Millipore.
MILLI-Q RES < SP, REPLACE
Q-GARD AND QUANTUM
The Milli-Q Water resistivity is <
set point.
Dispense water to eliminate any
trapped air in the System.
Replace the Q-Gard Pack and the
Quantum Cartridge.
Continued on next page