Ultrasonic milk analyser
Operation manual
11.04.19
98/110
pH Calibr OK
Which shows that the calibration procedure was completed successfully. The
calibrated device is ready for making measurements.
8. The device automatically passes in mode pH measuring.
9. Check the correctness of the calibration by measuring buffer solution 7.00.
9. pH Measuring.
After starting this menu the measurement is done in mode off line, i.e. the
analyser works only as a pH meter. The operator has to dip the probe in the
sample and on the display the following is shown:
pH measuring
x.xxxV
y.yy pH
Exit
Where:
x.xxx
– measured by the probe voltage
y.yyy
– measured probe’s pH
By pressing the button
Exit
, the operator may exit the program and to pass
towards another menu.
10. pH En/Disable
Serves for enabling/disabling the pH measuring during normal work of the
analyser - On line. After starting it the display shows:
pH Measuring
XXX
No OK Yes
Where:
XXX
is the current situation of the working mode. By pressing the buttons
below the corresponding inscriptions it could be changed, as
Yes
– means
that during normal work of the analyser – measuring the rest of the
parameters, pH will also be measured. If
No
is chosen, then pH is not
measured.
Содержание LACTOSCAN SA
Страница 12: ...Ultrasonic milk analyser Operation manual 11 04 19 12 110 Fig 3 Principle working scheme...
Страница 13: ...Ultrasonic milk analyser Operation manual 11 04 19 13 110 Fig 4 Connecting peripheral devices...
Страница 33: ...Ultrasonic milk analyser Operation manual 11 04 19 33 110 Fig 12 Peristaltic pump SL 16...
Страница 93: ...Ultrasonic milk analyser Operation manual 11 04 19 93 110 Fig 14 Placing the probe...