Ultrasonic milk analyser
Operation manual
11.04.19
97/110
3. Using the analyzer’s buttons enter the exact buffer value. For the probes
which are delivered with the analyzer, the isopotential point is always 7.00.
The following is displayed:
pH Calibr
Put Izopot buff
Buf=xx.xxx
4. The operator has to enter the buffer’s value, when the probe is in its
isopotential point and to press the button Enter.
After that the display shows:
pH Calibr
Put Izopot buff
Buf=xx.xxx
V=x.xxxV Set
Where
x.xxxxV
is the measured in the probe voltage.
5.Press the button
Set
when the readings stop moving.
6. Repeat the procedure with
Next buffer
and the display shows the
following:
Put filled with
Water glass
And Enter press
To continue
1. The operator has to place a glass filled with distilled water and to
press Enter, in order to clean the probe from previously used
calibration liquid. After the cleaning the analyzer is ready to start
working with the next calibration liquid.
2. Repeat the procedure with the
Next buffer
. The following message
appears on the display: