Ultrasonic milk analyser
Operation manual
11.04.19
61/110
What the operator sees during these passes is the difference in the speed of
sucking the sample. For that purpose, the high-fat sample has to be
preliminary heated up to
30С +- 3С.
To choose the mode, follow the sequence below:
Setup->Settings->Settings
Page2->Option
Select->SelPumpSpeed->Speed for Cal x
After which the display shows:
Calibr x
Pump Speed
Normal OK HiFat
By pressing the correspondent buttons the operator can choose the type of
measurement and to exit the menu.
When changing the type of measurement on a calibration is necessary a new
calibration of the device on the new speed to be done. When calibrating
measuring high fat sample, before starting the calibration procedure, the
operator has to choose from the menu:
Setup->Settings->Settings
Page2->Option
Select->SelPumpSpeed-> HFSpeed for Cal
By which the device passes in a mode of measuring high fat samples.
This calibration mode is active till the power supply of the device is switched
off i.e. it has to be always set if the device will be calibrated for high fat
measurement.
9.6. Working with embedded accumulator (option).
On customers’ request accumulator could be embedded in the milk analyser.
In this way it could work independently on the electrical network. In this case
the Power switch on the rear panel of the analyser serves for close / open the
electric circuit of the outer power supply. After closing the circuit, in order to
start the analyser, the operator has to press the Start button on the front
panel of the analyser.
Switching off the analyser could be done on one of the following ways:
Through the analyser – by switching the Power button on the rear panel of
the analyser.
Through the software – by using the option Shut Down from the Calibration
and Working mode menus.
Содержание LACTOSCAN SA
Страница 12: ...Ultrasonic milk analyser Operation manual 11 04 19 12 110 Fig 3 Principle working scheme...
Страница 13: ...Ultrasonic milk analyser Operation manual 11 04 19 13 110 Fig 4 Connecting peripheral devices...
Страница 33: ...Ultrasonic milk analyser Operation manual 11 04 19 33 110 Fig 12 Peristaltic pump SL 16...
Страница 93: ...Ultrasonic milk analyser Operation manual 11 04 19 93 110 Fig 14 Placing the probe...