2. Cleaning with acidic solution
Fill in the glass with warm (50-
60 С) acidic cleaning solution. Put it in the
recess of the analyser and start the command Cleaning from the main menu.
After finishing it, pour out the contaminated liquid.
3. Rinsing with water
Fill in the glass with water. Put it in the recess of the analyser and start
command “Cleaning” from the main menu. After finishing it pour out the
contaminated water. Now the device is ready for work.
Please, pay attention that, when the analysers gives a signal for need of
cleaning 15 min after the last measurement of real milk samples or 55 min.
after being powered and not used, cleaning is made ONLY with alkaline
solution in concentration 1-3%.
During the basic/final cleaning sequence is: alkaline solution – water – acidic
solution - water
IMPORTANT
THE MAIN REASON FOR MALFUNCTIONING OF THE
DEVICE IS THE BAD CLEANING OF THE SYSTEM AFTER
MAKING ANALYSIS.
In case of malfunction due to the bad cleaning of the
analyser your guarantee is not valid anymore and any
repair has to be paid.