Ultrasonic milk analyser
Operation manual
11.04.19
94/110
If you are working with the analyser regularly (each day) do not remove the
probe after work..
If you ‘ll not use the analyser more than 2 days, you must take out the probe
and to place the stopper back.
The pH probe must be stored separately as per the instructions of point 2.3
Appendix 3.
7. PH measuring.
Measuring pH is an additional feature of the analyser and is optional.
Remove the protective rubber cap of the pH electrode. Take care to handle it
appropriate in order not to be damaged. Use de-ionized or distilled water to
rinse the electrode before usage. Fill in the sample holder with milk, put it in
the recess of the analyser and dip the pH electrode into the milk sample,
ensuring complete dip of the electrode in the sample. Stir gently for
homogenization of the sample.
Measuring can be done in two modes:
Off line
by starting the menu
pH & Co Meter | Measuring
, when the
analyser works only as a pH meter.