2.6.2. Main power supply connection
FOR 230V VERSION, connect the cable CODE 50036 (left figure below) to connector (main power supply).
FOR 115V VERSION, connect the cable CODE 50391 (right figure below) to connector (main power supply).
Finally connect the plug of main power supply to socket: now it is possible to turn the DecalMATE ON.
2.6.3. Power supply connection (NO UPS mode)
See also the Label of the instrument that provides some important information.
Always communicate the S.N. (Serial Number) of the unit to Customer Support.
As separable power cord, it is strictly required to use power cord with wire size at least of
1.5mm² (equal to minimum AWG12). A size less of 1.5mm² (AWG12) cannot guarantee a
proper functionality of the unit and overheating of the power cord.
Characteristic: this is the easier way to supply DecalMATE, but it does NOT protect the unit from power
supply failure.
In fact, in case of power failure, user cannot remove samples from DecalMATE nor perform any
other operations.
For this reason, Milestone s.r.l. does NOT suggest this type of connection.
Make sure to inform the customer that in this way a power supply failure inhibits any operation with
DecalMATE.