BoneSTATION
MM093-002
– Operator Manual
31
3.3. Run only a decalcification program
If tissues have been previously fixed, it is possible to SKIP the fixation phase and proceed directly to the
DECALCIFICATION phase.
Put the DECALCIFIER metal holder, with glass containing decalcifier, on the FRONT MODULE if heated
decalcification is needed.
Then follow the instructions of chapter 3.2.2 on how to set program values and start a program.
Put the DECALCIFIER metal holder, with glass containing decalcifier, on the BACK MODULE if NOT heated
decalcification is needed.
Then follow the instructions of chapter 3.2.2 on how to set program values and start a program.
3.4. Run only a fixation program
If tissues need only to be fixed, it is possible to perform just a FIXATION phase.
Put the FIXATIVE metal holder, with glass containing fixative, on the BACK MODULE no heating of fixation
is needed (PRESOAKING).
Then follow the instructions of chapter 3.2.1 on how to set program values and start a program.
Put the FIXATIVE metal holder, with glass containing fixative, on the FRONT MODULE, if heated fixation is
required.
Then follow the instructions of chapter 3.2.1 on how to set program values and start a program.