Midtronics Inc. 7000 Monroe Street Willowbrook, IL 60527
www.midtronics.com
CELLTRON
MAX
•
35
•
Chapter 9: Troubleshooting
The troubleshooting tips in this section will help you resolve
most testing and printing problems. For problems with the
printer, digital temperature gun, or the PC software application,
refer to their manuals or call Midtronics Customer Service for
assistance. (See Patents, Limited Warranty, Service.)
Screen does not power on during testing (no text/graphics)
Check the connection to the jar.
The jar voltage might be too low (less than 1 volt) to test.
The analyzer’s battery pack might need to be recharged
or replaced.
Recharging the analyzer battery pack
Recharge the analyzer battery pack if:
The display does not turn on when you press the POWER
button.
The screen displays:
**Warning**
Internal Battery Low!
Replace Batteries Soon!
1. Insert the AC adapter plug into the connector as shown in
Display, keypad, and connections in Chapter 2.
2. Connect the power of the AC adapter to an AC outlet.
3. Periodically turn on the analyzer and check if the charge
level indicator is black. When the battery pack is fully
charged, disconnect the adapter from the analyzer and
the AC outlet.
NOTE: The maximum charge time is 3 to 4hours.
Do not overcharge.
Replacing the analyzer battery pack
If the screen does not power on after recharging, replace the
battery pack.
1. Press battery pack end tabs and pull battery pack.
2. Replace with charged battery pack.
If the problem persists, call Midtronics Customer Service.
(See Patents, Limited Warranty, Service.)
Probe tip is bent or stops retracting
To replace a damaged a probe tip:
1. Grasp the probe tip with pliers at the top of the sleeve.
Do not damage probe when removing
from sleeve.
Grasping the sleeve that encases the probe tip
can damage the tip.
2. Pull the tip straight out.
3. Grasp the replacement tip with the pliers and insert it into
the sleeve.
4. Push the probe tip into a soft surface, such as cardboard,
until it hits the bottom of the sleeve.
NOTE: To obtain replacement tips, contact
Midtronics Customer Service. (See Patents, Lim-
ited Warranty, Service.)
Test Failure
If the analyzer fails to advance to the next jar count, try to
retest.
Ensure clamps are connected and LEDs are off.
Test results do not print or print incorrectly
Check that the printer is on
Check that the tester IR transmitter are aligned
Check printer batteries
Flourescent lights can affect IR transmission. Remove the
tester from any flourescent lights and re-transmit.