220
⚫
UART serial decode
The measured signal word length is 8bit; parity bit, none; baud rate, 19.2kb/s, hexadecimal; trigger mode as
data bit:55; follow the steps as below:
(1) Tap S1 to open the decode channel, and click S1 again to open the bus configuration menu;
(2) Select the bus type as “UART”, click “Ch1”, “Idle High”, “Parity None”, “8bit”, “19.20kb/s”, display
“hexadecimal”, then close menu;
(3) Open the trigger mode setting menu, click “Data”, enter 55 manually, and press “enter” to confirm;
(4) Adjusting the threshold level according to the amplitude level of signal may make the signal to be stably
triggered. The UART trigger graphic interface is shown in Figure 12-8:
Method 1: Click
configuration information to open the decode channel threshold level adjustment
box, and drag the adjustment box upward and downward to adjust the threshold level.
Содержание Smart Series
Страница 1: ......
Страница 2: ...Version Info Version Date Remarks V2 0 2019 10 ...
Страница 25: ...Chapter 2 Quick Start Guide of Oscilloscope 11 2 3 Side Panel Figure 2 2 Side Panel ...
Страница 27: ...Chapter 2 Quick Start Guide of Oscilloscope 13 2 5 Front Panel Figure 2 4 Front Panel of Tablet Oscilloscope ...
Страница 30: ...16 Figure 2 5 Oscilloscope Interface Display ...
Страница 51: ...Chapter 3 Operate the Oscilloscope with Buttons 37 Figure 3 1 Button Areas ...
Страница 61: ...Chapter 4 Horizontal System 47 Figure 4 1 Horizontal system ...
Страница 82: ...68 Figure 5 4 Channel Open Close and Switching ...
Страница 88: ...74 Figure 5 7 DC Coupling Figure 5 8 AC Coupling ...
Страница 126: ...112 Figure 6 21 Trigger Level Adjustment Trigger pulse width time can be set as 8ns 10s ...
Страница 150: ...136 Figure 7 2 Pull up Menu Figure 7 3 All Measurements ...
Страница 159: ...Chapter 7 Analysis System 145 Figure 7 8 Frequency Meter Measurement Menu Open Figure 7 9 Frequency Meter Measurement ...
Страница 161: ...Chapter 7 Analysis System 147 Figure 7 10 Cursor Measurement Description ...
Страница 168: ...154 Figure 8 1 Screen Capture Please refer to 13 6 Picture View for details on viewing pictures ...
Страница 176: ...162 Figure 8 5 Delete Reference Files CSV files CSV file structure ...
Страница 190: ...176 Figure 9 4 Spectrum Amplitude as V Hz Adjust FFT waveforms Waveform position ...
Страница 236: ...222 Figure 12 8 UART Graphic Interface UART graphic interface description 1 Trigger position 2 Trigger type ...
Страница 324: ...310 Figure 14 6 Android APP Figure 14 7 Successful Connection of Android APP ...
Страница 336: ...322 Figure 15 3 Decode Functions Installed ...
Страница 353: ...Annex 339 Optional accessories 1 Oscilloscope suitcase handbag 2 Battery 3 Leather carry strap 4 High voltage probe ...