Chapter 6 Trigger System
107
2) Trigger condition and pulse width time setting
Time restrictions that can set in the trigger condition: <, >,
=
, ≠.
⚫
Smaller than the time value (<)
For example, for positive pulse, if it is set as T<80ns, the trigger will happen stably only when the pulse
width is smaller than 80ns (Figure 6-15 Trigger Time T<80ns).
Figure 6-15 Trigger Time T<80ns
⚫
Greater than the time value (>)
For example, for positive pulse, if it is set as T>80ns, the trigger will happen stably only when the pulse
width is greater than 80ns (Figure 6-16 Trigger Time T>80ns).
Figure 6-16 Trigger Time T>80ns
⚫
Equal to the time value (=)
For example, for positive pulse, if it is set as T=80ns, the trigger will happen stably only when the pulse
width is equal to 80ns (Figure 6-17 Trigger Time T=80ns).
Figure 6-17 Trigger Time T=80ns
⚫
Not equal to the time value (≠)
For example, for positive pulse, if it is set as T≠80ns, the trigger will happen stably only when the pulse
width is not equal to 80ns (Figure 6-18 Trigger Time T≠80ns).
Trigger
Trigger
Trigger