Chapter 12 Maintenance
8. Set your current source to the value that you want to use for the upper reference point and
wait 3-5 seconds until the value displayed by the DeviceConfig has stabilised. Ensure that the
required value is entered for the upper reference point before you press the button to adopt
the upper reference point for the relevant universal input (see ""Calibration" tab " on page
100). You can also use the "Set all upper" button to set the upper reference points for both of
the universal inputs at the same time. However, each of the universal inputs requires its own
current source for this purpose. The current sources must all be set to the same value during
this process.
9. Press the "Save" button to transfer the calculated offset and slope values to the myDa-
talog2Ex (see ""Calibration" tab " on page 100).
10. The "Print" button is used to generate the calibration protocol that must always be created to
ensure traceability. After pressing the "Print" button, a dialogue window will appear in which
additional information can be added. The report is printed once this window is closed by
selecting "OK". The standard printer is always used for this process without any further que-
ries. A PDF writer can also be used.
Rev. 02
101
Содержание 305001
Страница 2: ......
Страница 8: ......
Страница 9: ...Chapter 2 Declaration of conformity Chapter 2 Declaration of conformity Rev 02 9 ...
Страница 10: ......
Страница 11: ...Chapter 3 Ex certification Chapter 3 Ex certification Rev 02 11 ...
Страница 12: ...12 Rev 02 ...
Страница 13: ...Chapter 3 Ex certification Rev 02 13 ...
Страница 14: ......
Страница 26: ......
Страница 38: ......
Страница 64: ......
Страница 84: ......
Страница 106: ......
Страница 108: ......
Страница 126: ......