Page 32
UF AV5500 Area-Velocity Flow Meter
FIELD TROUBLESHOOTING
The UF AV5500 uses an ultrasonic level sensor to determine flow area and an ultrasonic Doppler sensor
to measure flow velocity.
The QZ02L sensor combines both sensors in one housing.
An optional configuration uses the PZ15-LP “down-looking” level sensor and a QZ02L-B velocity
sensor.
To troubleshoot the UF AV5500, verify correct operation of level and velocity measurements separately.
Note: Selecting “
Restore
Defaults
” in the
SPECIAL FUNCTION
menu will return the instrument to “as-
shipped” factory settings.
LEVEL (QZ02L SENSOR)
SYMPTOMS
FAULTS
SOLUTIONS
EC
percent at zero
- very turbulent flow
- very aerated flow
- relocate sensor or use PZ15-LP
- sensor not level
- level sensor with “Bullseye” level
- sediment/dirt/grease build-up on
sensor
- clean sensor with liquid soap
Level Inaccurate
- sensor not mounted at bottom of
pipe/channel
- set a “Lvl Offset” in Calibration
menu
VELOCITY (QZ02L SENSOR)
SYMPTOMS
FAULTS
SOLUTIONS
- No velocity reading
- grease/sediment on sensor
- improper hook-up
- not enough suspended solids or
aeration in water.
- clean sensor with detergent
- check sensor connections
- drop Alka-Seltzer tab into flow
stream to create bubbles to verify
lack of air in pipe.
- water too clean for Doppler
- perform bucket test to prove
Doppler works.
- Velocity reading too High/Low
- poor velocity profile in
channel/pipe
- Cal Constant changed
- release sensor per suggestions in
sensor location section
- change Cal Constant in
Calibration menu
Содержание UF AV5500
Страница 2: ...Note This page has been left blank intentionally...
Страница 28: ...Page 28 UF AV5500 Area Velocity Flow Meter...
Страница 34: ...Page 34 UF AV5500 Area Velocity Flow Meter Resistance vs Temperature...
Страница 40: ...Page 40 UF AV5500 Area Velocity Flow Meter SS PIPE MOUNTING BAND OPTION VSJ...
Страница 61: ...Page 61 UF AV5500 Area Velocity Flow Meter Dual Non Contacting Sensor Configuration...