Page 23
UF AV5500 Area-Velocity Flow Meter
the calibration menus.
Press
✓
to return to
Menu Selections
.
SIMULATION
Exercises the 4-20mA (0-5V) outputs, digital display and control relays.
Test
Select
Maximum
and press
✓
to simulate maximum Flow, Level and
Velocity and to output 20mA (5V) to the analog channels.
Select
Minimum
and press
✓
to simulate minimum Flow, Level and
Velocity and to output 4mA (0V) to the analog channels.
To simulate an intermediate Flow, Level and Velocity set
Test
to
Actual
and then enter a value for the
Level
and
Velocity
. The Flow
calculation, analog outputs and control relays will respond to the
simulated values.