UF AV5500 Area-Velocity Flow Meter
INSTALLATION
– SENSOR LOCATION
For the most accurate flow measurement possible, careful consideration should be made to the
placement of the sensor in relation to flow disturbances. In general, the best accuracy will occur where
flow is evenly distributed across the channel/pipe and free of turbulence.
Specific installation considerations are listed and discussed in more detail below.
1.
Open Discharges or Pipe/Channel Outfalls
When the QZ02 sensor is to be mounted in front (upstream) of an open discharge or pipe/channel
outfall, the sensor should be placed at least 5 times the maximum head level in front of the
outfall: