Page 14
UF AV5500 Area-Velocity Flow Meter
CALIBRATION (cont.)
4mA
Set the 4mA value for the level analog output, in units
configured in the Units/Mode menu.
Velocity
Min
Only shown when Mode = Velocity in the Units/Mode menu.
When measured velocity is less than the Min Velocity, the
reading on the LCD display and output signals will report 0.
20mA
Set the 20mA value for the velocity analog output, in units
configured in the Units/Mode menu.
4mA
Set the 4mA value for the velocity analog output, in units
configured in the Units/Mode menu.
Flow
Min
Only shown when Mode = Flow in the Units/Mode menu.
When measured flow is less than the Min Flow, the reading on
the LCD display and output signals will report 0.
20mA
Set the 20mA value for the flow analog output, in units
configured in the Units/Mode menu.
4mA
Set the 4mA value for the flow analog output, in units
configured in the Units/Mode menu.
Lvl Offset
For the QZ02L sensor only. Set to 0.00 when sensor is
mounted at the bottom of a channel/pipe. When the sensor is
mounted above the floor, in order to avoid having the sensor
become coated with debris, enter the distance between the
floor and the bottom of the sensor.
- - Ca l i b r a t i o n - - - - - - -
Ra n g e
Ma x
1 8 0 . 0 0 i n
Mi n
8 . 0 0 i n
L e v e l
Ma x
1 8 0 . 0 0 i n
Mi n
1 . 0 0 i n
2 0 mA
1 8 0 . 0 0 i n
4 mA
0 . 0 0 i n
Ve l o c i t y
2 0 mA
1 0 . 0 0 f t / s
4 mA
0 . 0 0 f t / s
Fl o w
Mi n
0 . 0 0 g a l / m
2 0 mA
5 0 0 . 0 0 g a l / m
4 mA
0 . 0 0 g a l / m
L v l Of f s e t
0 . 0 0 i n
Si g n a l Cu t o f f
1 0 %
Da mp i n g
Ve l o c i t y
1 0 %
L e v e l
1 0 %
L OE Ti me
3 0 s e c
Ca l Co n s t a n t
1 . 0 0 0
Содержание UF AV5500
Страница 2: ...Note This page has been left blank intentionally...
Страница 28: ...Page 28 UF AV5500 Area Velocity Flow Meter...
Страница 34: ...Page 34 UF AV5500 Area Velocity Flow Meter Resistance vs Temperature...
Страница 40: ...Page 40 UF AV5500 Area Velocity Flow Meter SS PIPE MOUNTING BAND OPTION VSJ...
Страница 61: ...Page 61 UF AV5500 Area Velocity Flow Meter Dual Non Contacting Sensor Configuration...