background image

                                                                                          Optical Strain Gages |

 os3150 os3155

 

C o p y r i g h t   ©   2 0 1 3   M i c r o n   O p t i c s ,   I n c .  

 

 

 

     

 

 

 

 

o s 3 1 5 0   o s 3 1 5 5   I n s t a l l a t i o n   G u i d e ,   R e v i s i o n   C       

 

 

 

 

   

3

 

 

the sensor. The coarser surface is required for proper adhesion of the waterproof seal and protective paint 
topcoat. Wipe the surface clean with a compatible solvent and inspect the surface to ensure that all surface 
defects have been removed. 
 

 

Thoroughly clean surface using a clean cloth saturated with a compatible solvent 

 
 

Gage Installation: 

 

 

Position the gage on the surface to be tested. Align the center axis of the gage with the strain axis to be 
measured. The gage may be held into position with adhesive tape. When applying tape, do not cover weld 
pockets. Figures 3 and 4 show the gage axis and weld sequence. The gage is welded in place at four 
primary weld pockets 1 thru 4. Four secondary Weld Pockets A thru D further secure the gage. 
 

 

Tack gage in place using a single spot weld in weld pocket 1. 
 

 

Perform a single weld in each of the subsequent weld pockets 2, 3, 4. These welds should be performed in 
the order indicated. 
 

 

Add 4 additional welds to each weld pocket. The welds should be equally spaced within the weld pocket as 
shown in Figure 5. 
 

 

It is important that the gage be securely welded in place with at least 5 welds per pocket. If the gage is not 
securely fastened to the test specimen, measurement errors will result.  

 

 

 

Figure 3

 

– os3150 Weld Locations 

 
 

 

 

Figure 4

 

– os3155 Weld Locations 

 
 

Содержание os3150

Страница 1: ...I n s t a l l a t i o n G u i d e R e v C Optical Strain Gage os3150 os3155 Micron Optics Inc 1852 Century Place NE Atlanta GA 30345 USA phone 404 325 0005 fax 404 325 4082 www micronoptics com...

Страница 2: ...on FBG In addition accurate measurement requires that the Stain gage temperature compensator and test specimen surface be at or near the same temperature A layer of insulation over the sensor will hel...

Страница 3: ...The gage may be held into position with adhesive tape When applying tape do not cover weld pockets Figures 3 and 4 show the gage axis and weld sequence The gage is welded in place at four primary weld...

Страница 4: ...0 x 250mm should be prepared to SSPC SP10 NACE 2 prior to installing the gage as shown in Figure 6 It is important that this area be free from all paint rust scale oil and grease This will allow the g...

Страница 5: ...2 square pieces of STOPAQ Paste 1 x 1 x 1 8 inch 25 x 25 x 3mm Place one piece under the cable near each end of the gage as shown in Figure 7 Figure 7 STOPAQ Paste Under Cable Cut 4 rectangular piece...

Страница 6: ...Position STOPAQ over the gage and center as shown in Figure 10 Starting from the center and working toward the edges gently press the STOPAQ in place forcing out any air Firmly press STOPAQ in place a...

Страница 7: ...of the patch Paint over the primer with a compatible topcoat such as Carboxane 2000 Figure 12 Primed and Painted Accurate measurement of strain requires that the Stain gage temperature compensator an...

Страница 8: ...o s 3 1 5 5 I n s t a l l a t i o n G u i d e R e v i s i o n C 8 Figure 13 os315X Protective Cover Insulation Kit The installed cover is shown in Figure 14 Figure 14 Covered Sensor The drawing for t...

Страница 9: ...Optical Strain Gages os3150 os3155 C o p y r i g h t 2 0 1 3 M i c r o n O p t i c s I n c o s 3 1 5 0 o s 3 1 5 5 I n s t a l l a t i o n G u i d e R e v i s i o n C 9 Figure 15 Cover Drawing...

Страница 10: ...e connector or splice on the end of the sensor cable also needs to be protected from the elements If the cable end is not protected the cable could become a path for moisture to enter the sensor One m...

Отзывы: