background image

                                                                                          Optical Strain Gages |

 os3150 os3155

 

C o p y r i g h t   ©   2 0 1 3   M i c r o n   O p t i c s ,   I n c .  

 

 

 

     

 

 

 

 

o s 3 1 5 0   o s 3 1 5 5   I n s t a l l a t i o n   G u i d e ,   R e v i s i o n   C       

 

 

 

 

   

6

 

 

 

 

 

Figure 9

 

– STOPAQ Paste Detail 

 

 

Cut a piece of STOPAQ WrappingBand EZ  4  x 7 ½ inch [100 x 190mm] 
 

 

Position STOPAQ over the gage and center as shown in Figure 10. Starting from the center and working 
toward the edges, gently press the STOPAQ in place, forcing out any air. Firmly press STOPAQ in place 
around the edges of gage. 
 

 

 

Figure 10

 

– STOPAQ EZ Patch 

 

 

Firmly press and kneed the STOPAQ around the perimeter of the gage to form a tight seal with the base 
metal. Pay special attention to the area around the cable to ensure that it is sealed as shown in Figure 11. 
 

Содержание os3150

Страница 1: ...I n s t a l l a t i o n G u i d e R e v C Optical Strain Gage os3150 os3155 Micron Optics Inc 1852 Century Place NE Atlanta GA 30345 USA phone 404 325 0005 fax 404 325 4082 www micronoptics com...

Страница 2: ...on FBG In addition accurate measurement requires that the Stain gage temperature compensator and test specimen surface be at or near the same temperature A layer of insulation over the sensor will hel...

Страница 3: ...The gage may be held into position with adhesive tape When applying tape do not cover weld pockets Figures 3 and 4 show the gage axis and weld sequence The gage is welded in place at four primary weld...

Страница 4: ...0 x 250mm should be prepared to SSPC SP10 NACE 2 prior to installing the gage as shown in Figure 6 It is important that this area be free from all paint rust scale oil and grease This will allow the g...

Страница 5: ...2 square pieces of STOPAQ Paste 1 x 1 x 1 8 inch 25 x 25 x 3mm Place one piece under the cable near each end of the gage as shown in Figure 7 Figure 7 STOPAQ Paste Under Cable Cut 4 rectangular piece...

Страница 6: ...Position STOPAQ over the gage and center as shown in Figure 10 Starting from the center and working toward the edges gently press the STOPAQ in place forcing out any air Firmly press STOPAQ in place a...

Страница 7: ...of the patch Paint over the primer with a compatible topcoat such as Carboxane 2000 Figure 12 Primed and Painted Accurate measurement of strain requires that the Stain gage temperature compensator an...

Страница 8: ...o s 3 1 5 5 I n s t a l l a t i o n G u i d e R e v i s i o n C 8 Figure 13 os315X Protective Cover Insulation Kit The installed cover is shown in Figure 14 Figure 14 Covered Sensor The drawing for t...

Страница 9: ...Optical Strain Gages os3150 os3155 C o p y r i g h t 2 0 1 3 M i c r o n O p t i c s I n c o s 3 1 5 0 o s 3 1 5 5 I n s t a l l a t i o n G u i d e R e v i s i o n C 9 Figure 15 Cover Drawing...

Страница 10: ...e connector or splice on the end of the sensor cable also needs to be protected from the elements If the cable end is not protected the cable could become a path for moisture to enter the sensor One m...

Отзывы: