![Microm HM 355 S Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/microm/hm-355-s/hm-355-s_instruction-manual_1787359069.webp)
Rotary Microtome HM 355 S
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
386 200 - English
5-3
REPLACEMENT WORK
Replacing the fuses:
The two fuses of the instrument are placed on the
rear side of the instrument beside the main switch
(fig. 73.1).
•
To replace the two fuses, turn off the main
switch of the instrument and unplug it.
•
Then open the cover (fig. 73.2) by means of a
flat screwdriver.
•
Pull out the fuse holders, each of which is
marked with an arrow, and replace the fuses
with new ones.
Rating of fuses:
For power requirements 220 – 240 V:
- 2 fuses 1,6 AT slow-blow
For power requirements 100 – 120 V:
- 2 fuses 3,2 AT slow-blow
Replacing the fuse for the motor supply:
The fuse for the motor supply (fig. 73.4) is placed
on the rear side of the instrument on the right side
of the mains plug unit and secures the supply
voltage of the cutting drive.
•
To replace the two fuses, turn off the power
switch of the instrument and unplug the
instrument.
•
Press the fuse holder cap by means of a
screwdriver and turn it to the right side.
•
Remove the fuse and replace it with a new
one.
•
Insert the cap with the fuse into the fuse
holder and lock the bayonet catch.
Rating of fuse:
- 1 fuse 5,0 AT, slow-blow (cat. no. 326350)
Fig. 73