Rotary Microtome HM 355 S
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
386 200 - English
3-4
SETTING SECTION THICKNESS AND
TRIMMING THICKNESS
•
The required section and trimming
thicknesses are set by means of the control
knob (fig. 16.1), which is placed on the side
part of the operating panel.
•
To choose between section thickness and
trimming thickness, press the control knob (fig.
16.1). The selected thickness range is then
marked with a rim when in the operating mode
NORMAL (fig. 17.1).
Note:
Switching over from trim sectioning to
fine sectioning via the operating knob
(fig. 16.1) can also serve to reduce the
motorized cutting speed (see part "Automatic
speed reduction" 3-9).
Note:
Whenever the instrument is turned on,
"FINE" sectioning is automatically
selected and shown accordingly.
FINE = pre-selected section thickness
TRIM = pre-selected trimming thickness
The graduation of the section thicknesses is
divided into five ranges:
Range: Graduation:
up to 5 µm
0,5 µm
from 5 µm to 20 µm
1 µm
from 20 µm to 30 µm
2 µm
from 30 µm to 60 µm
5 µm
from 60 µm to 100 µm
10 µm
The graduation of the trimming thicknesses is
divided into four ranges:
Range: Graduation:
up to 10 µm
5 µm
from 10 µm to 100 µm
10 µm
from 100 µm to 200 µm
20 µm
from 200 µm to 500 µm
50 µm
Start
Stop
Retraction
MEMO
MENU
Fig. 16
Fig. 17
1
2
FINE= 9
TRIM= 30
SUM = 0
R : ON
0
1