Rotationsmikrotom HM 355 S
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
387840 - Deutsch
•
Connect the plug of the foot pedal cable into
the outlet (fig. 3.6) and tighten the swivel nut.
Note:
If the foot pedal is not connected, there
is the operating mode emergency stop
(see part 3-6-7). This way, the
handwheel brake is activated and the cutting drive
motor cannot be started. Always connect the foot
pedal!
•
Connect the power cord to the power socket
(fig. 3.4) on the rear side of the instrument.
•
Turn on the power switch (fig. 3.5).
•
Then, the specimen holder moves to the rear
end position.
•
This movement is always carried out when the
instrument is turned on.
•
This way, the instrument is calibrated.
•
The insert for the two fuses is placed beside
the power switch.
•
The fuse for the motor supply (fig. 4.4) is
placed on the right side of the mains plug unit.
•
Replacement of the instrument fuses, please
see part 5-3.