![Microm HM 355 S Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/microm/hm-355-s/hm-355-s_instruction-manual_1787359026.webp)
Rotary Microtome HM 355 S
MICROM International GmbH
Robert-Bosch-Str. 49
D- 69190 Walldorf
386 200 - English
3-3-4 EMERGENCY
STOP
To quickly eliminate danger, the microtome has
two "emergency stop" devices.
Caution:
In case danger arises from cutting
drive, push the "emergency stop"!
The
hand emergency stop button
(fig. 9.1) is
placed on the right side of the microtome below
the handwheel.
•
Press this red button to stop immediately the
motorized cutting drive.
•
STOP is shown on the display of the
operating panel (fig. 10.1), if the "emergency
stop" is activated.
•
Continue sectioning, pull out the red button.
The cutting drive can be started again.
The
second emergency stop device
is
integrated into the foot pedal.
•
Vigorously step on the foot pedal to
immediately stop the motorized cutting drive.
•
This emergency stop device is activated as
long as the foot pedal is being stepped on.
•
STOP is shown on the display (fig. 10.1) of
the operating panel, if the "emergency stop" is
activated.
•
Continue sectioning, release the foot pedal.
The cutting drive can be started again.
Fig. 9
Fig. 10
1