4-2
HOW TO AVOID MALFUNCTIONS
Specimen preparation:
•
When preparing specimens, be sure that a
suitable embedding medium, fixation,
dehydration and infiltration time are chosen.
Specimen temperature:
•
Sectioning is carried out at ambient
temperature (excluding frozen sections).
•
If the temperature is too high, the paraffin
softens. Therefore, avoid heating paraffin
specimens by direct exposure to sunlight or
other near sources of heat.
Tightening the clamping screws:
•
Tighten all clamping screws and clamping
levers on the knife carrier, specimen holder
and specimen orientation.
Selection of the knife:
•
Carefully select the required knife material
and profile.
Adjustment of the knife:
•
Take care to adjust the proper clearance
angle of the knife: 5 - 7° for glass knives. For
diamond knives ask the knife manufacturer.
•
Select a clearance angle adjustment of 5 - 15°
according to the facet angle. Typically, adjust
an angle of 10 - 12 °.
•
Take care to adjust knife height.
Cutting speed:
Take care to select proper cutting speed.
•
General Rule: The harder the material, the
slower the cutting speed!
Trimming:
•
Take care in bringing knife and specimen
together.