UPS Universal Power Supply – Model # 200-006R / 200-004R - Instructions For Use 
For use only with 

Microline Cautery Instruments and Microline PowerPack™ Footswitch 

Device Description 

The Model # 200-006R / 200-004R

 

Universal Power Supply (UPS) is a non-sterile, reusable, AC powered unit intended to be used only 

with Microline Cautery Instruments and the PowerPack Footswitch. The UPS is intended to be connected to AC power via a grounded 
hospital grade power cord and features an on/off switch and green power-on LED indicator.  Depending on the device used, up to three 
operating heat ranges are available from the UPS with corresponding tones indicating the power level of the instruments heating 
element. A low frequency tone indicates the use of low heat. An interrupted low tone indicates the use of variable heat, and a higher 
frequency continuous tone indicates the use of high heat levels. If the instrument does not require the use of the PowerPack 
Footswitch, the UPS will supply power at the variable heat level only, unless a multi-heat range device is used. The UPS may be placed 
on a flat non-sterile surface adjacent to the sterile field, or may be suspended from a nearby IV pole. If suspended from an IV pole, the 
UPS may be rotated to face the user. 

Intended Use 

The UPS is intended for use only with Microline Cautery Instruments and the PowerPack Footswitch for the simultaneous cutting and 
cauterization of soft tissue during surgery. 
1. Connect grounded, Hospital Grade power cord to the UPS AC connector receptacle. 
2. Plug the power cord into a grounded Hospital Grade receptacle.  
3. Align the key portion of the Microline Cautery Instrument connector with the UPS connector receptacle and insert.  
4. If the PowerPack Footswitch is to be used, plug the Footswitch connector into the Footswitch connector receptacle of the UPS. 
5. Turn on the UPS. 
6. Refer to the appropriate Microline Cautery Instrument and PowerPack Footswitch (if applicable) Instructions For Use for device 
specific use instructions and pre-check requirements. 
7. Model # 200-006R:

 

When the instrument heating element is activated, the UPS emits a tone that will last until the heating element is 

deactivated.  The amount of heat delivered by the instrument heater may be adjusted by pressing the “+” or “-” push pads on the right 
side of the UPS. The numerical setting is shown on the LED bar graph above the push pads.  Increasing the LED’s numerical value 
setting increases the amount of heat delivered by the instrument when using a single-heat range instrument or the variable range when 
using the optional PowerPack Footswitch accessory or multi-heat range instrument.  Generally, lower heat ranges increase the sealing 
capabilities and increase the time required to divide tissue.  Higher heat ranges decrease the time to divide and may decrease seal 
integrity. 
8. Model # 200-004R:

  

When the instrument heating element is activated, the UPS (Universal Power Supply) emits a tone that will last 

until the heating element is deactivated.  The amount of heat delivered by the device heater may be adjusted by varying the knob 
setting on the UPS (Universal Power Supply).  Increasing the knob’s numerical value setting increases the amount of heat delivered by 
the device when using a single-range device or the variable range when using the optional PowerPack Footswitch accessory or multi-
range device. Generally, lower heat ranges increase the sealing capabilities and increase the time required to divide tissue.  Higher 
heat ranges decrease the time to divide and may decrease seal integrity. 
9. At the end of the surgical procedure, turn the UPS off and disconnect the Microline Cautery Instrument and if applicable, the 
PowerPack Footswitch. 
Handling, Storage And Cleaning 
Do not drop. Store in a cool, dry place. Avoid prolonged exposure to extreme temperatures. Exterior of the UPS may be cleaned using 
a soft cloth moistened with a solution of water and a mild detergent or disinfectant. Do not immerse the UPS in liquids. 
Recommended Environmental Conditions 
 

 

Transport 

Storage 

Operating 

 

-15ºC to 50º 

15ºC to 30º 

15ºC to 30º 

 

25 to 95 

30 to 85 

45 to 75 

 

700hPa to 1060hPa 

700hPa to 1060hPa

 

700hPa to 1060hPa

 

 
There are no unusual risks associated with the proper disposal of this equipment. Follow any local regulations regarding proper 
disposal of used electrical equipment. 
 

Model # 200-006R 
 

Universal Power Supply 

1-  Hanger 
2-  Volume LED Indicators 
3-  Volume Control Buttons 
4-  Footswitch Connector Receptacle and Indicator 
5-  AC Connector Receptacle 
6-  Power On LED Indicator 
7-  On/Off Switch 
8-  Cautery Instrument Connector Receptacle and indicator 
9-  Power Level Control Buttons 
10- Power Level LED Indicators 

 

 

 
Model # 200-004R 
 

Universal Power Supply 

Non Sterile 

1-  Power On LED indicator 
2-  Hanger 
3-  Footswitch Connector and Indicator 
4-  AC Connector 
5-  On/Off Switch 
6-  Cautery Instrument Connector and Indicator 

 

 
 
Note

: A power cord for North America (U.S. and Canada only) is supplied with the UPS. For other locales, ensure that the appropriate 

power cord and connector are used, per National Guidelines (i.e. HAR, 250 VAC, 2A, 3 x 0.75 mm2, minimum). 

Содержание 200-004R

Страница 1: ...mount of heat delivered by the instrument when using a single heat range instrument or the variable range when using the optional PowerPack Footswitch accessory or multi heat range instrument Generall...

Страница 2: ...under one or more of US patents 7 033 351 6 908 463 6 860 880 6 695 837 6 626 901 Patents Pending All Rights Reserved Copyright 2011 Microline Surgical Inc All rights reserved Bloc d alimentation uni...

Страница 3: ...minimum Mises en garde et avertissements Lors de l utilisation aux tats Unis utiliser uniquement avec le cordon d alimentation de qualit h pitaux fourni raccord une prise de qualit h pitaux Utiliser u...

Страница 4: ...chiave del connettore dello strumento di cauterizzazione Microline con la presa del connettore dell UPS e inserire 4 Se si intende usare la pedaliera PowerPack inserire il connettore della pedaliera...

Страница 5: ...uso continuo Si raccomanda un ciclo di servizio di circa cinque 5 secondi di servizio e dieci 10 secondi di pausa Se l UPS viene usato ininterrottamente o il connettore dello strumento cortocircuitato...

Страница 6: ...b der Drucktasten angezeigt Durch Erh hen der Einstellung des numerischen Werts wird die von dem Instrument abgegebene Hitze erh ht wenn ein Instrument mit einfachem Hitzebereich oder der variable Ber...

Страница 7: ...1 A max Ausgang max 60 W Spannung ohne Last 9 V Gleichstrom Einhaltung der Normen Das Netzger t erf llt die Anforderungen von IEC60601 1 UL 2601 1 und CAN CSA C22 2 Nr 601 1 M90 f r Ger te der Klasse...

Страница 8: ...ntervenci n quir rgica apagar el UPS posici n Off y desconectar el instrumento cauterizador de Microline y si corresponde el Conmutador de pedal PowerPack Manipulaci n almacenamiento y limpieza No dej...

Страница 9: ...veau van het verwarmingselement op het instrument aangeven Bij gebruik van laag vermogen lage temperatuur wordt een laag geluidssignaal gegeven Een onderbroken laag geluidssignaal geeft aan dat er var...

Страница 10: ...te zijn met het anatomisch gebied waarin ze de ingreep wensen te verrichten Raadpleeg bij gebruik van het UltraSlim of SC Forceps instrument de gebruiksaanwijzing voor informatie over de inspectie v r...

Страница 11: ...006R N r instrumentets varmeelement aktiveres afgiver den universale str mforsyningsenhed en tone der vedvarer indtil varmeelementet deaktiveres Den m ngde varme der leveres af instrumentets varmeelem...

Страница 12: ...ks effekt 60W Hvilesp nding 9V j vnstr m Overensstemmelse med standarder Den universale str mforsyningsenhed er i overensstemmelse med kravene i IEC60601 1 UL 2601 1 og CAN CSA C22 2 Nr 601 1 M90 for...

Страница 13: ...nesteisiin Ymp rist olosuhteet Kuljetus S ilytys K ytt 15 50 C 15 30 C 15 30 C 25 95 30 85 45 75 700 1060 hPa 700 1060 hPa 700 1060 hPa T m n laitteen asianmukaiseen h vitt miseen ei liity mit n ep t...

Страница 14: ...rst ende intraven sstativ Dersom UPS enheten henger fra et intraven sstativ kan den roteres slik at den vender mot brukeren Tiltenkt bruk UPS enheten er beregnet til bruk kun med Microlines kauterise...

Страница 15: ...eller erstattes av brukeren Merknad UPS enheten er ikke beregnet p kontinuerlig bruk En typisk driftssyklus p omtrent fem 5 sekunder p ti 10 sekunder av anbefales Dersom UPS enheten brukes kontinuerl...

Страница 16: ...r nda tills v rmeelementet inaktiveras M ngden v rme som levereras av instrumentets v rmeelement kan justeras genom att man trycker p eller p n taggregatets h gra sida Den numeriska inst llningen ange...

Страница 17: ...regatet uppfyller kraven enligt IEC60601 1 UL 2601 1 och CAN CSA C22 2 nr 601 1 M90 f r utrustning av klass I typ CF och uppfyller kraven p elektromagnetisk kompatibilitet enligt IEC60601 1 2 Undvik a...

Страница 18: ...auteriza o Microline e se aplic vel o Interruptor a Pedal PowerPack Manuseamento Armazenamento e Limpeza N o deixe cair Armazene em lugar fresco e seco Evite exposi o prolongada a temperaturas extrema...

Страница 19: ...i a t i provozn teplotn rozsahy s odpov daj c mi t ny indikuj c mi v kon topn ho t lesa n stroj T n s n zkou frekvenc indikuje pou it n zk ho ru P eru ovan t n indikuje pou v n p eru ovan ho ru a nep...

Страница 20: ...im nebo SC Forceps postupujte podle kontroly p ed pou it m uveden m v n vodu k pou it V tomto za zen nejsou dn sou stky na kter ch m e u ivatel prov d t servis Pozn mka UZN nen ur en k nep etr it mu p...

Страница 21: ...ry trwa b dzie do czasu jego dezaktywacji Ilo dostarczonego ciep a z grzejnika instrumentu mo na ustawia przyciskami z symbolami or po prawej stronie zasilacza UPS Ustawienia numeryczne pokazane s na...

Страница 22: ...mocy UPS Wska e na to brak sygna u Przed dalszym post powaniem nale y ustali przyczyn problemu Je eli problemu nie mo na rozwi za lub przywr ci mocy po kilku minutach ch odzenia nale y skontaktowa si...

Страница 23: ...8 200 004R UPS UPS PowerPack 9 UPS off Microline PowerPack UPS UPS 15 C 50 C 15 C 30 C 15 C 30 C 25 C 95 30 C 85 45 C 75 700hPa 1060hPa 700hPa 1060hPa 700hPa 1060hPa 200 006R 1 2 LED 3 4 5 6 LED 7 On...

Страница 24: ...0035 D 7 033 351 6 908 463 6 860 880 6 695 837 6 626 901 Copyright 2011 Microline Surgical Inc UPS 200 006R 200 004R Microline Microline PowerPack Footswitch UPS 200 006R 200 004R AC Microline UPS LED...

Страница 25: ...UPS UPS Microline Bovie Microline UltraSlim SC Forceps UPS 5 10 UPS UPS Microline Surgical UPS 100 240 AC 50 60 100 60 9 DC UPS IEC60601 1 UL 2601 1 and CAN CSA C22 2 No 601 1 M90 Class I CF CF IEC60...

Страница 26: ...s kijelz se Egyszeres h m rs klet tartom ny k sz l kn l vagy pedig az opcion lis PowerPack l bkapcsol tartoz k eset n v ltoztathat tartom nyn l illetve t bbsz r s h m rs klet tartom ny k sz l kn l a L...

Страница 27: ...bel l a t pell t s nem ll vissza vegye fel a kapcsolatot a Microline Surgical c ggel a k sz l k jav t s ra vagy cser j re vonatkoz utas t sok rt Figyelmeztet s Tilos gy l kony anyag jelenl t ben haszn...

Страница 28: ...UPS UPS 15 C 50 15 C 30 15 C 30 25 95 30 85 45 75 700 1060 700 1060 700 1060 200 006R 1 2 3 4 5 6 7 8 9 10 200 004R 1 2 3 4 5 6 A UPS HAR 250 VAC 2A 3 x 0 75 mm2 minimum Microline PowerPack UPS UPS U...

Страница 29: ...S IEC60601 1 UL 2601 1 CAN CSA C22 2 No 601 1 M90 I CF t CF IEC60601 1 2 UPS Microline Surgical Inc UPS 1 Microline Microline Surgical Inc CF IFU0035 Rev D 7 033 351 6 908 463 6 860 880 6 695 837 6 62...

Страница 30: ...A G 4 1 6 9 19 9 9AB 3 5 6 1 5 78 9 9 L N P 7 200 006R 9 B E G H 01 0 4 2 7 7 1 4 5 1 95 5 4 C E 1 01 0 1 23 4 4 54 B H 54 0 89 AB 1 56 1 4 8 200 004R 8 9 A 9 D F G 3 56 7 5 5B B 0 7 B H I B J K L 78...

Страница 31: ...lim SC 6 9 B D 3 3 3 5 3 5 10 34 7 9 4 7 G I OP 4 4 4 78 4 4BC 4 4 G7 H 3 7 D D G K DM 4 2 4 5 0 04 7 9 4 100 240 50 60 100 60 9 0 1 0 14 IEC60601 1 1UL 2601 1 CAN CSA C22 2 601 1 M90 I 12 CF 1 8 IEC6...

Страница 32: ...daha d k s aral klar cihaz n kapatma yetene ini ve doku divizyon s resini artt r r Y ksek s aral klar divizyon s resini k salt r ve kapat lan dokunun b t nl n azaltabilir 8 Model No 200 004R Aletin s...

Страница 33: ...anestetikler bulundu u yerlerde kullanmay n Aleti atmadan nce daima ba lant lar n kesin GK tekrar kullan labilir Elektrik Gereksinimleri Giri 100 240 VAC 50 60 Hz 100W Maksimum k 60W Y ks z Voltaj 9V...

Отзывы: