5
NC 150
EN
13. Guarantee
This device is covered by a
5 year guarantee
from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
Transport costs and risks of transport.
Damage caused by incorrect application or non-compliance
with the instructions for use.
Damage caused by leaking batteries.
Damage caused by accident or misuse.
Packaging/storage material and instructions for use.
Regular checks and maintenance (calibration).
Accessories and wearing parts: Batteries, Probe cover
(optional).
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
14. Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial technical
inspection is recommended for professional users. Please
observe the applicable disposal regulations.
15. www.microlife.com
Detailed user information about our thermometers and blood pressure
monitors as well as services can be found at www.microlife.com.
Type:
Non Contact Thermometer NC 150
Measurement
range:
Body mode: 34.0 - 43 °C / 93.2 - 109.4 °F
Object mode: 0.1 - 99.9 °C / 32.2 - 211.8 °F
Resolution:
0.1 °C / °F
Measurement
accuracy:
Laboratory:
0.2 °C, 36.0
39.0 °C /
0.4 °F, 96.8
102.2 °F
Display:
L
iquid
C
rystal
D
isplay, 4 digits plus special icons
Acoustic:
The unit is turned ON and ready for the measure-
ment: 1 short beep.
Complete the measurement: 1 long beep (1 sec.) if
the reading is less than 37.5 °C / 99.5 °F, 10 short
«beep» sounds, if the reading is equal to or greater
than 37.5 °C / 99.5 °F.
System error or malfunction: 3 short «bi» sounds.
Memory:
30 readings recall in the memory mode with a
record of both time and date.
Backlight:
The display light will be GREEN for 1 second,
when the unit is turned ON.
The display light will be GREEN for 5 seconds,
when a measurement is completed with a reading
less than 37.5 °C / 99.5 °F.
The display light will be RED for 5 seconds, when
a measurement is completed with a reading equal
to or higher than 37.5 °C / 99.5 °F.
Operating
conditions:
Body mode: 15 - 40.0 °C / 59 - 104.0 °F
Object mode: 5 - 40.0 °C / 41 - 104.0 °F
Storage
conditions:
-25 - +55 °C / -13 - +131 °F
15 - 95 % relative maximum humidity
Automatic
Switch-off:
Approx. 1 minute after last measurement has been
taken.
Battery:
2 x 1.5 V alkaline batteries; size AAA
Battery
lifetime:
approx. 2000 measurements
(using new batteries)
Dimensions:
141.1 x 43.3 x 36.9 mm
Weight:
90 g (with batteries), 67 g (w/o batteries)
IP Class:
IP21
Reference to
standards:
ASTM E1965; IEC 60601-1; IEC 60601-1-2 (EMC);
IEC 60601-1-11
Expected
service life:
5 years or 12000 measurements