15
EN
aponorm
®
Basis
Using rechargeable batteries
You can also operate this instrument using rechargeable batteries.
10. Using a Mains Adapter
You can operate this instrument using the Microlife mains adapter
(DC 6V, 600mA).
When the mains adapter is connected, no battery current is
consumed.
)
Please use only type «NiMH» reusable batteries!
)
The batteries must be removed and recharged, if the battery
symbol (battery flat) appears! They must not remain inside
the instrument, as they may become damaged (total
discharge as a result of low use of the instrument, even
when switched off).
)
Always remove the rechargeable batteries, if you do not
intend to use the instrument for a week or more!
)
The batteries can NOT be charged in the blood pressure
monitor! Recharge these batteries in an external charger
and observe the information regarding charging, care and
durability!
)
Use only the Microlife mains adapter available as an orig-
inal accessory appropriate for your supply voltage, e.g. the
«Microlife 230V adapter».
)
Ensure that neither mains adapter nor cable are not
damaged.
1. Plug the adapter cable into the mains adapter socket
5
in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
You can operate this instrument using the aponorm
®
by microlife
adapter (DC 6V, 600mA).
Use only the aponorm
®
by microlife mains adapter avai-
lable as an original accessory appropriate for your supply
voltage, e.g. the «aponorm
®
by microlife 230V adapter».
7.
11. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «
ERR 3
», is displayed.
* Please consult your doctor, if this or any other problem occurs
repeatedly.
Error
Description Potential cause and remedy
«
ERR 1
» Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
«
ERR 2
» Error signal
During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be gener-
ated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate
and no result can therefore be displayed.
Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«
ERR 6
» MAM Mode There were too many errors during the
measurement in MAM mode, making it
impossible to obtain a final result. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
«
HI
»
Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«
LO
»
Pulse too
low
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
)
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
8.
* Please consult your doctor, if this or any other problem occurs
repeatedly.
«
HI
»
Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«
LO
»
Pulse too
low
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
)
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
Содержание aponorm Basis
Страница 45: ...45 RU aponorm Basis AQ AR AS II 3 x x 3 1 x x 1 AN 1...
Страница 48: ...48 aponorm by 12 O 1 x x O Q x x x x x x x 9 30 C 2 Microlife x aponorm by microlife...
Страница 50: ...50 aponorm by 2 5 c 10 109193 6 095 679 96 96...
Страница 51: ...51 RU aponorm Basis...
Страница 95: ...95 GR aponorm Basis 9 AN 10 AQ AR AS X 9 4 3 1 2 5 3 4 5 x x 3 cm 1 in x x 6 ON OFF 1 7 8 11 II 12 1 ON OFF...
Страница 96: ...96 aponorm by 6 M AT M AP M 17 17 M CL M CL 200 Full M 9 AO AO 1 7 2 3 2 4 1 5V AA 30 4 6 7 3 IV 5...
Страница 98: ...98 aponorm by 11 ERR 3 ERR 1 ERR 2 ERR 3 ERR 5 ERR 6 MAM MAM 8 12 HI 300 mmHg 200 5 LO 40 1 x x x x 9...
Страница 101: ...101 GR aponorm Basis...