You must enter
Two-Phase Flow Low Limit
in g/cm
³
, even if you configured another unit
for density measurement.
The default value for
Two-Phase Flow Low Limit
is 0.0 g/cm
³
. The range is 0.0 to
3.0 g/cm
³
.
2.
Set
Two-Phase Flow High Limit
to the highest density value that is considered normal in
your process.
Values above this will cause the transmitter to post Alert A105 (Two-Phase Flow).
Tip
To reduce the occurrence of two-phase flow alerts that are not significant to your process, set
Two-Phase Flow High Limit
slightly above your expected highest process density.
You must enter
Two-Phase Flow High Limit
in g/cm
³
, even if you configured another
unit for density measurement.
The default value for
Two-Phase Flow Low Limit
is 3.0 g/cm
³
. The range is 0.0 to
3.0 g/cm
³
.
3.
Set
Two-Phase Flow Timeout
to the number of seconds that the transmitter will wait for
a two-phase flow condition to clear before posting the alert.
The default value for
Two-Phase Flow Timeout
is 0.0 seconds, meaning that the alert
will be posted immediately. The range is 0.0 to 60.0 seconds.
Detecting and reporting two-phase flow
Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of
process control issues. By configuring the two-phase flow parameters appropriately for
your application, you can detect process conditions that require correction.
Tip
To decrease the occurrence of two-phase flow alerts, lower
Two-Phase Flow Low Limit
or raise
Two-Phase
Flow High Limit
.
A two-phase flow condition occurs whenever the measured density goes below
Two-Phase
Flow Low Limit
or above
Two-Phase Flow High Limit
. If this occurs:
•
A two-phase flow alert is posted to the active alert log.
•
Line density is held at its last
pre‐alert
value for the number of seconds configured in
Two-Phase Flow Timeout
.
If the two-phase flow condition clears before
Two-Phase Flow Timeout
expires:
•
Line density reverts to actual process density.
•
The two-phase flow alert is deactivated, but remains in the active alert log until it is
acknowledged.
If the two-phase flow condition does not clear before
Two-Phase Flow Timeout
expires, line
density reverts to actual process density, but the two-phase flow alert remains active.
Configure process measurement
Configuration and Use Manual
27
Содержание CDM100M
Страница 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Страница 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Страница 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Страница 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Страница 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Страница 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Страница 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Страница 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Страница 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...
Страница 218: ...Concentration measurement matrices derived variables and process variables 212 Micro Motion Compact Density Meters CDM ...
Страница 219: ...Concentration measurement matrices derived variables and process variables Configuration and Use Manual 213 ...