4.
In Step 1, ensure that the setting of
Derived Variable
matches the derived variable
used by your matrix. If it does not, change it as required and click
Apply
.
Important
If you change the setting of
Derived Variable
, all existing concentration matrices will be deleted
from transmitter memory. Verify the setting of
Derived Variable
before continuing.
5.
Load one or more matrices.
a. In Step 2, set
Matrix Being Configured
to the location (slot) to which the matrix will
be loaded.
b. Click
Load Matrix from a File
, navigate to the matrix file on your computer, and load
it.
c. Repeat until all required matrices are loaded.
6.
(Optional) Set the density and temperature units to the units you want to use for
measurement.
Related information
Matrix switching
Derived variables and calculated process variables
Configure
Set reference temperature values for specific gravity using
ProLink III
When
Derived Variable
is set to
Specific Gravity
, you must set the reference temperature to be
used for density measurement and the reference temperature of water, and then verify
the density of water at the configured reference temperature. The two reference
temperature values affect specific gravity measurement.
Additionally, the two reference temperature values affect any concentration process
variable that is calculated by equation, rather than by matrix, because the equations are
based on specific gravity.
Typically, the two reference temperature values are the same, but this is not required.
Restriction
If
Derived Variable
is not set to
Specific Gravity
, do not change any of these values. These are set by the
active concentration matrix.
Procedure
1.
Choose
Device Tools
>
Configuration
>
Process Measurement
>
Concentration Measurement
.
The Concentration Measurement window is displayed. It is organized into steps that
allow you to perform several different setup and configuration tasks. For this task,
you will not use all the steps.
Configure process measurement
60
Micro Motion
®
Compact Density Meters (CDM)
Содержание CDM100M
Страница 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Страница 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Страница 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Страница 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Страница 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Страница 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Страница 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Страница 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Страница 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...
Страница 218: ...Concentration measurement matrices derived variables and process variables 212 Micro Motion Compact Density Meters CDM ...
Страница 219: ...Concentration measurement matrices derived variables and process variables Configuration and Use Manual 213 ...