5.
Replace the safety spacer and end-cap.
6.
Power up the meter.
3.4
Set the HART lock
If you plan to use a HART connection to configure the meter, you can lock out all other
HART masters. If you do this, other HART masters will be able to read data from the meter
but will not be able to write data to the meter.
Restrictions
•
This feature is available only when you are using the Field Communicator or AMS.
•
This feature requires HART 7.
Procedure
1.
Choose
Configure
>
Manual Setup
>
Security
>
Lock/Unlock Device
.
2.
If you are locking the meter, set
Lock Option
as desired.
Option
Description
Permanent
Only the current HART master can make changes to the device. The device will
remain locked until manually unlocked by a HART master. The HART master can
also change
Lock Option
to
Temporary
.
Temporary
Only the current HART master can make changes to the device. The device will
remain locked until manually unlocked by a HART master, or a power-cycle or
device reset is performed. The HART master can also change
Lock Option
to
Perma-
nent
.
Lock All
No HART masters are allowed to make changes to the configuration. Before
changing
Lock Option
to
Permanent
or
Temporary
, the device must be unlocked. Any
HART master can be used to unlock the device.
Postrequisites
To avoid confusion or difficulties at a later date, ensure that the meter is unlocked after
you have completed your tasks.
3.5
Restore the factory configuration
Display
Not available
ProLink III
Device Tools
>
Configuration Transfer
>
Restore Factory Configuration
Field Communicator
Service Tools
>
Maintenance
>
Reset/Restore
>
Restore Factory Configuration
Introduction to configuration and commissioning
18
Micro Motion
®
Compact Density Meters (CDM)
Содержание CDM100M
Страница 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Страница 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Страница 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Страница 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Страница 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Страница 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Страница 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Страница 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Страница 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...
Страница 218: ...Concentration measurement matrices derived variables and process variables 212 Micro Motion Compact Density Meters CDM ...
Страница 219: ...Concentration measurement matrices derived variables and process variables Configuration and Use Manual 213 ...