Alert num-
ber
Alert title
Possible cause
Recommended actions
A030
Incorrect Board Type The loaded software is not compati-
ble with the programmed board
type.
•
Contact Micro Motion.
A033
Insufficient Pickoff
Signal
The signal from the sensor pick-
off(s) is insufficient. This suggests
that the sensor tubes or vibrating
elements are not vibrating. This
alert often occurs in conjunction
with Alert 102.
•
Check for foreign material in the proc-
ess gas or fluid, coating, or other proc-
ess problems.
•
Check for fluid separation by monitor-
ing the density value and comparing
the results against expected density val-
ues.
•
Ensure that the sensor orientation is ap-
propriate for your application. Settling
from a two-phase or three-phase fluid
can cause this alert.
A037
Sensor Check Failed
Known Density Verification failed.
•
Check the subtest results and perform
the recommended actions.
•
Retry the test.
•
Contact Micro Motion.
A038
Time Period Signal
Out of Range
The time period signal is outside the
limits for the sensor type.
•
Check your process conditions against
the values reported by the device.
A100
mA Output 1 Satura-
ted
The calculated mA output value is
outside the configured range.
•
Check the settings of
Upper Range Value
and
Lower Range Value
. See
•
Check process conditions. Actual condi-
tions may be outside the normal condi-
tions for which the output is config-
ured.
•
Check for foreign material in the proc-
ess gas or fluid, coating, or other proc-
ess problems.
•
Verify that the measurement units are
configured correctly for your applica-
tion.
A101
mA Output 1 Fixed
The HART address is set to a non-
zero value, or the mA output is con-
figured to send a constant value.
•
Check whether the output is in loop test
mode. If it is, unfix the output.
•
Exit mA output trim, if applicable.
•
Check the HART address. If the HART
address is non-zero, you may need to
change the setting of
mA Output Action
(
Loop Current Mode
).
•
Check whether the output has been set
to a constant value via digital communi-
cation.
Troubleshooting
146
Micro Motion
®
Compact Density Meters (CDM)
Содержание CDM100M
Страница 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Страница 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Страница 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Страница 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Страница 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Страница 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Страница 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Страница 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Страница 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...
Страница 218: ...Concentration measurement matrices derived variables and process variables 212 Micro Motion Compact Density Meters CDM ...
Страница 219: ...Concentration measurement matrices derived variables and process variables Configuration and Use Manual 213 ...